Toll-Free in Philippines: 1 800 8908 8267 | 1 800 1441 0877
International+65 6415 5353

DRST Series Communication Programmer

DRST-CM

Communication Programmer for DRST Series | DRST-CM

Communication Programmer for DRST Series

PHP10,905.90 DRST-CM

CE
  • Programming Display for DRST Series
  • Modbus Protocol Interface Over RS485
  • Monitor Process Value From the Built-In Display
  • High 2.5 kV Isolation to Host Unit
  • Shielded RJ45 Connector on Top
USB, RS232, RS485 and Ethernet Data Acquisition Systems - View related products

Description

Application
• The DRST-CM detachable display adds Modbus RTU RS-485 serial communications to all current and future DRST Series units.
• The unit converts a wide array of sensors and analog device signals measured by the system like uni- and bipolar mA and voltage signals, potentiometer, Lin. R, RTD and TC, to a Modbus communication line signal.
• All individual unit operating parameters can easily and quickly be configured by using the Modbus communication or by using the front display menu.
• The easily readable DRST-CM display can be used to read the process signal, simulate the output signal, indicate sensor errors and internal device errors.

Technical Characteristics
• DRST-CM has full DRSL-Display functionality for unit programming, process signal monitoring and diagnostics handling.
• Modbus RTU protocol is supported using a serial RS-485 communication wiring.
• Multi-drop half-duplex connection via shielded RJ45 connector.
• High safe galvanic isolation of 2.5 kVAC between the serial wiring and the connected system DRST units.
• Modbus parameters such as address, baud rate, stop bit(s), and parity bit are configured from the DRST-CM display, which also stores parameters.

Mounting/Installation/Programming
• Mounting in Zone 2 / Div 2. This DRST-CM can be moved from one device to another. The individual system DRST unit configuration of the first device can be saved and downloaded to subsequent devices.
• Programmed parameters can be protected by a user defined password.
• When mounted on devices that are installed upside down, a menu item allows the display on DRST-CM to be rotated 180o and the up/down buttons to switch function.

Specifications
Environmental Conditions
Operating Temperature: -20 to 60°C
Storage Temperature: -20 to 85°C
Relative Humidity: < 95% RH (non-cond.)
Protection Degree: IP20
Installation In: Pollution degree 2 and measurement/overvoltage cat. II
Mechanical Specifications
Dimensions (HxWxD): 73.2 x 23.3 x 26.5 mm
Dimensions (HxWxD) w/4000/9000 Unit: 109 x 23.5 x 131 mm
Weight Approx: 100 g
Connection: RJ45 - shielded
Common Specifications
Supply
Max. Required Power: = 0.15 W
Isolation Voltage
Isolation Voltage, Test/Working: 2.5 kVac/250 Vac reinforced isolation
Response Time
Response Time: < 20 ms
Signal/Noise Ratio: > 60 dB
Update Rate: > 50 Hz
Extended EMC Immunity: NAMUR NE21, A Criterion, Burst: No loss of communication
Signal Type: RS485 half duplex
Serial Protocol: Modbus RTU
Modbus Mode: RTU - slave
Devices on an RS485 Line:Up to 32 (without a repeater)
Data Rates, Baud: 2400, 4800, 9600, 19200, 38400, 57600, 115200
Automatic Baudrate Detection: Yes - can be configured ON or OFF
Parity: Even, Odd, None
Stop Bit(s): 1 or 2
Digital Addressing: 1 to 247
Response Delay:0 to 1000 ms
Observed Authority Requirements
EMC: 2014/30/EU
LVD: 2014/35/EU
EAC TR-CU 020/2011
Approvals
DNV-GL Marine: Stand. f. Certific. No. 2.4
ATEX 2014/34/EU: DEKRA 13ATEX0098 X
IECEx: DEK 13.0026 X
FM: 0003049132-C
UL: UL 61010-1
(Specify Model Number)
Part Number/Desc.
Qty
Description
DRST-CM
PHP10,905.90
Availability: 1 week
 
DRST Series Communication Programmer
All prices on this site are shown in Philippine Pesos.
We noticed you are from the US. Visit your local site for regional offers and live support.