Intrinsically Safe Temperature and Humidity Data Logger
|
Click for larger image. |
|
|
$780.00
OM-CP-RHTEMP1000IS-A2
|
|
Certified Intrinsically Safe
|
|
SGS Certified to meet FM3600 and FM3610 standards
|
|
Software Battery Life Indicator
|
|
316 Stainless Steel Enclosure
|
|
Small, Sleek Design
|
|
Fast Response Time
|
|
CE Compliant
|
View related products - Data Loggers
|
| |
|
|
|
The OM-CP-RHTEMP1000IS-A2 is an Intrinsically Safe humidity and temperature data logger. It carries hazardous location, intrinsically safe certification in accordance with the latest issue of FM3600, FM3610. This certification makes the device ideal for EtO Sterilization, environmental studies, and numerous other hostile environment applications.
The device can be started and stopped directly from your computer and its small, sleek size allows it to fit almost anywhere. The OM-CP-RHTEMP1000IS-A2 makes data retrieval quick and easy. Simply insert the device into the OM-CP-IFC400 or the OM-CP-IFC406 USB docking station (sold separately) and our user-friendly software does the rest The small size of the OM-CP-RHTEMP1000IS-A2, allows it to be placed in remote or hard to reach locations. The device now features a battery life indicator allowing users to easily identify low voltage and change batteries when required. This feature maximizes the use of your battery life and is a cost saving efficiency.
The OM-CP Data Logger Software is a powerful, analytic tool. Data can be viewed in graphical or tabular formats and summary and statistics views are available for further analysis.
Specifications
TEMPERATURE
|
Temperature Sensor
|
Resistance Temperature Detector (RTD)
|
Temperature Range
|
-40 °C to +80 °C (-40 °F to +176 °F)
|
Temperature Resolution
|
0.01 °C (0.018 °F)
|
Calibrated Accuracy
|
±0.5 °C (0 °C to ±50 °C) ±0.9 °F (32 °F to 122 °F)
|
HUMIDITY
|
Humidity Sensor
|
Capacitive Polymer
|
Humidity Range
|
0 %RH to 100 %RH (non-condensing)
|
Humidity Resolution
|
0.1 %RH
|
Calibrated Accuracy
|
±3.0 %RH maximum
|
Specified Accuracy Range
|
25 %RH to 75 %RH +20 °C to +40 °C (68 °F to 104 °F) Hysteresis Error 1 % typical, 3 % Maximum
|
GENERAL
|
Memory
|
32,768 Readings
|
Start Modes
|
Software programmable immediate start or delay start, up to 24 months in advance
|
Real Time Recording
|
May be used with PC to monitor and record data in real time
|
Reading Rate
|
1 reading every second up to 1 reading every 24 hours
|
Calibration
|
Digital calibration through software
|
Calibration Date
|
Automatically recorded within device
|
Battery Type
|
Tadiran TL-2150 3.6V lithium battery included, user replaceable in a non-hazardous location
|
Battery Life
|
2 years typical at 15 minute reading rate
|
Data Format
|
Date and time stamped °C, °F, K, °R %RH, mg/mL, Dew Point
|
Time Accuracy
|
16 seconds / month -40 °C to 80 °C
|
Computer Interface
|
OM-CP-IFC400 or OM-CP-IFC406
|
Operating System Compatibility
|
Windows XP SP3/7/8/10
|
Software Compatibility
|
Standard Software version 4.2.19.1 or later Secure Software version 4.2.18.0 or later
|
Operating Environment
|
-40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH (case properly sealed)
|
Dimensions
|
1.67 in x 0.97 in dia. (42.3 mm x 24.6 mm dia.)
|
Weight
|
2.4 oz (68 g)
|
Enclosure
|
316 Stainless Steel/Radel
|
Approvals
|
CE FM Class 3600, Rev. Jan 2018 FM Class 3610, Rev. Jan 2018 CAN/CSA-C22.2 No. 60079-0:15 CAN/CSA-C22.2 No. 60079-11:14
|
|
| |
|
|
|
|
|
† All dollar amounts on this site are shown in Singapore currency.
|
|
|
|
|
| |
|
|