X-Series - Multi Channel Voltage Loggers
|
Click for larger image. |
|
|
$1,980.00
OM-CP-OCTVOLT-32V-A2
|
|
Real-time Operation
|
|
Programmable Engineering Units
|
|
24-bit Readings Provide High Resolution
|
|
Up to 4 Hz Reading Rate
|
|
Compact Size
|
|
External Power or User Replaceable Battery
|
|
Pushbutton or Programmable Start Time
|
|
Configurable Channel Names
|
|
NIST Traceable Calibration Certificate Included
|
View related products - Data Loggers
|
| |
|
|
|
The new OM-CP-VoltX Series consists of 4, 8, 12 and 16-channel, battery powered, DC voltage data logger with a reading rate of up to 4 Hz. X-Series devices ship with a standard USB-A to Micro USB cable for use with free OMCP PC based software. The X-series has a high memory capacity and increased download speed compared to previous generations of product.
This easy-to-use device features an 18 month typical battery life, user replaceable battery and programmable engineering units. The storage medium is non-volatile solid state memory, providing maximum data security even if the battery becomes discharged. To maximize memory capacity, users can enable or disable channels. For easy identification, each channel can be named with up to a ten digit title.
Using the OM-CP Data Logger Software, starting, stopping and downloading from the OM-CP-VoltX Series data loggers are simple and easy. Graphical, tabular and statistical data is provided for analysis and data can be viewed in multiple units, using the Engineering Units function. The data can also be automatically exported to Excel® for further calculations.
Specifications
MEASUREMENT
|
Input Connection
|
Removable screw terminals
|
Model
|
160 mV |
3.2 V |
32 V |
Voltage Range
|
±160 mV |
±3.2 V |
-8 V to +32 V |
Voltage Resolution
|
0.02 mV |
0.0004 mV |
0.004 mV |
Calibrated Accuracy
|
±0.032 mV |
±0.0032 V |
±0.02 V |
Input Impedance
|
1 MΩ |
125 kΩ |
125 kΩ |
Maximum Voltage Between Inputs to Ground
|
+3 V |
+50 V |
+50 V |
Analog Conversion Time
|
150 ms
|
Frequency Rejection
|
60+50 Hz
|
Temperature Coefficient
|
25 ppm/℃
|
Specified Accuracy Range
|
Nominal range @ 25 ℃
|
Engineering Units
|
Native Measurement units can be scaled to display measurement units of another type. This is useful when monitoring current outputs from different types of sensors such as temperature, CO2, flow rate and more.
|
GENERAL
|
Memory (All channels enabled)
|
4-channel: 698,709 readings per channel 8-channel: 349,354 readings per channel 12-channel: 232,903 readings per channel 16-channel: 174,677 readings per channel
|
Start Modes
|
Software programmable immediate start or delay start, up to 6 months in advance
|
Real Time Recording
|
May be used with PC to monitor and record data in real time
|
LEDs
|
1 per channel and 2 status LEDs
|
Reading Rate
|
4 Hz up to 1 reading every 24 hours
|
Calibration
|
Digital calibration through software
|
Calibration Date
|
Automatically recorded within device
|
Battery Type
|
9 V lithium included, user replaceable
|
Battery Life
|
18 months typical
|
Data Format
|
Date and time stamped V, mV, µV engineering units specified through software
|
Time Accuracy
|
±1 minute/month
|
Computer Interface
|
USB-A to micro USB cable (included); 460,800 baud
|
Operating System Compatibility
|
Windows XP SP3/7/8/10
|
Software Compatibility
|
Standard Software version 4.2.19.0 or later Secure Software version 4.2.18.0 or later
|
Operating Environment
|
-20 ℃ to +60 ℃(-4 ℉ to + 140 ℉ ), 0 %RH to 95 %RH non-condensing
|
Dimensions
|
4-channel: 2.70 in × 7.25 in × 1.22 in (65.6 mm × 184.2 mm × 31.0 mm) 8-channel: 2.70 in × 7.25 in × 1.22 in (65.6 mm × 184.2 mm × 31.0 mm) 12-channel: 2.70 in × 7.25 in × 1.68 in (65.6 mm × 184.2 mm × 42.7 mm) 16-channel: 2.70 in × 7.25 in × 2.14 in (65.6 mm × 184.2 mm × 54.4 mm)
|
Weight
|
4-channel: 13 oz (368 g) 8-channel: 13 oz (368 g) 12-channel:20 oz (580 g) 16-channel: 28 oz (800 g)
|
Enclosure
|
Black anodized aluminum
|
Approvals
|
CE, ROHS
|
|
| |
|
|
|
|
|
† All dollar amounts on this site are shown in Singapore currency.
|
|
|
|
|
| |
|
|
Keywords:voltage datalogger,dataloggers,voltage logger,om-cp-volt
|
Back to top
|
|